ORIGINAL ARTICLE
Year : 2013  |  Volume : 56  |  Issue : 2  |  Page : 114-119

Comparison of platelet counts by sysmex XE 2100 and LH-750 with the international flow reference method in thrombocytopenic patients


Department of Hematology, P. D. Hinduja Hospital & Medical Research Centre, Mumbai, Maharashtra, India

Correspondence Address:
Tina Dadu
Department of Hematology, P. D. Hinduja Hospital & Medical Research Centre, Veer Savarkar Marg, Mahim, Mumbai - 400 016, Maharashtra
India
Login to access the Email id

Source of Support: National Health & Education Society (work done at P.D. Hinduja Hospital & Medical Research Centre, Mumbai), Conflict of Interest: None


DOI: 10.4103/0377-4929.118701

Rights and Permissions

Background: There are several methods for counting platelets, of which the international flow reference method (IRM) is considered to be the gold standard. We compared the platelet count given by this method to the count given by automated analyzers using other methods, such as optical fluorescence and impedance. Aims: The aim of this study is to compare the platelet counts obtained by Sysmex XE 2100 by Impedance (Sysmex-I), optical florescence (Sysmex-O) and reported (Sysmex-R) based on the switching algorithm and LH-750 by Impedance (LH-750) with the IRM in thrombocytopenic blood samples. To calculate the sensitivity, specificity, positive predictive value (PPV) and negative predictive value (NPV) of various technologies at the clinically relevant transfusion thresholds of 10 × 10 9 /l and 20 × 10 9 /l. Materials and Methods: A total of 118 blood samples with platelet count of <50 × 10 9 /l were selected for the study. Platelet counts of all samples were analyzed by all methods using the Sysmex analyzer, LH-750 and IRM in parallel within 6 h of collection. Statistical Analysis Used: Pearson correlation, bland Altman analysis, sensitivity and specificity, PPV and NPV. Results and Conclusions: Sysmex-R had the least Bias and 95% limits of agreement (95%LA) range and thus correlated best with IRM values. LH-750 had a higher Bias compared to Sysmex-O and Sysmex-R, but a strikingly similar 95% LA ensures similar results in all three methods. In fact, in the oncology subset, it had the narrowest 95% LA, which made it the best performer in this subgroup. Of the three Sysmex results, Sysmex-I had the highest bias, widest 95% LA and highest potential risk of over transfusion. Hence, Sysmex-R and LH-750 were found to be reliable tools for estimation of platelet count in thrombocytopenic patients.


[FULL TEXT] [PDF]*
Print this article     Email this article
 Next article
 Previous article
 Table of Contents

 Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
 Citation Manager
 Access Statistics
 Reader Comments
 Email Alert *
 Add to My List *
 * Requires registration (Free)
 

 Article Access Statistics
    Viewed5662    
    Printed136    
    Emailed5    
    PDF Downloaded221    
    Comments [Add]    
    Cited by others 3    

Recommend this journal